DAkkS calibration of resistance thermometers (Pt100, Pt25, etc.)

Calibration of resistance thermometers (temperature sensors such as Pt100, Pt25, etc.) and indicators with DAkkS calibration certificate is done at slim ITS-90 fixed points. It is also possible to make calibrations according to the comparison method in liquid baths or calibration furnaces according to the guideline DKD-R 5-1 “Calibration of resistance thermometers”.

Guideline DKD-R 5-1 “Calibration of resistance thermometers” from -196 °C to 962 °C
Fixed point calibration from -189 °C to 962 °C
Turnaround time is five working days
Temperature probes from a length of 300mm can be calibrated
Temperature probes and indicators

Uncertainty of measurement during DAkkS calibration

Temperature rangeProcedureUncertainty of measurement
-189,3442 °CDKD-R 5-1:2018 Argon triple point4,0 mK
-38,8344 °CDKD-R 5-1:2018 Mercury triple point2,5 mK
0,01 °CDKD-R 5-1:2018 Water triple point1,5 mK
29,7646 °CDKD-R 5-1:2018 Gallium melting point2,0 mK
156,5985 °CDKD-R 5-1:2018 Indium melting point4,5 mK
231,928 °CDKD-R 5-1:2018 Tin freezing point3,0 mK
231,928 °CDKD-R 5-1:2018 Tin melting point6,0 mK
419,527 °CDKD-R 5-1:2018 Zinc freezing point3,0 mK
419,527 °CDKD-R 5-1:2018 Zinc melting point10 mK
660,323 °CDKD-R 5-1:2018 Aluminium freezing point5,0 mK
660,323 °CDKD-R 5-1:2018 Aluminium melting point18 mK
961,78 °CDKD-R 5-1:2018 Silver freezing point10 mK
Calibration at temperature fixed points.
Temperature rangeProcedureUncertainty of measurement
-189,3442 °C to 0 °CDKD-R 5-1:2018 fixed points: Ar, Hg, TPW5,5 mK
-38,8344 °C to 29,7646 °CDKD-R 5-1:2018 fixed points: Hg, TPW, Ga3,5 mK
0 °C to 29,7646 °CDKD-R 5-1:2018 fixed points: TPW, Ga2,5 mK
0 °C to 156,5985 °CDKD-R 5-1:2018 fixed points: TPW, In melting points5,5 mK
0 °C to 231,928 °CDKD-R 5-1:2018 fixed points: TPW, In, Sn
melting points
7,0 mK
0 °C to 419,527 °CDKD-R 5-1:2018 fixed points: TPW, Sn, Zn
freezing points
4,5 mK
0 °C to 419,527 °CDKD-R 5-1:2018 fixed points: TPW, Sn, Zn
melting points
12 mK
0 °C to 660,323 °CDKD-R 5-1:2018 fixed points: TPW, Sn, Zn, Al freezing points7,0 mK
0 °C to 660,323 °CDKD-R 5-1:2018 fixed points: TPW, Sn, Zn, Al melting points20 mK
0 °C to 961,78 °CDKD-R 5-1:2018 fixed points: TPW, Sn, Zn, Al, Ag freezing points12 mK
Calibration at temperature fixed points with determination of characteristic curve. The measurement uncertainty refers to the characteristic curve in the specified range.
Temperature rangeProcedureUncertainty of measurement
-196 °CDKD-R 5-1:2018 in liquid nitrogen with compensation block15 mK
-180 °C to -80 °CDKD-R 5-1:2018 in cryostat60 mK
-80 °C to -60°CDKD-R 5-1:2018 in calibration furnace or liquid bath15 mK
> -60 °C to 90 °CDKD-R 5-1:2018 in calibration furnace or liquid bath10 mK
> 90 °C to 200 °CDKD-R 5-1:2018 in calibration furnace, liquid- or solid bath30 mK
> 200 °C to 500 °CDKD-R 5-1:2018 in calibration furnace, liquid- or solid bath40 mK
> 500 °C to 660 °CDKD-R 5-1:2018 in calibration furnace, liquid- or solid bath50 mK
> 660 °C to 962 °CDKD-R 5-1:2018 in calibration furnace, liquid- or solid bath0,15 K
Comparison with standard resistance thermometers.

Calibration at temperature fixed points

Klasmeier’s calibration service calibrates not only SPRTs but also resistance thermometers (Pt100, Pt25, Pt1000, working standard thermometers, etc.) using ITS-90 temperature fixed points for you. Each resistance thermometer is annealed at the highest temperature of use before calibration and the annealing drift is determined using the water triple point. Self-heating errors and hydrostatic pressure effects are investigated and taken into account during calibration. During each calibration, the hysteresis and the ITS-90 characteristic curve are also checked. The DAkkS calibration certificate is internationally accepted and can also be used abroad.

Short immersion depth from 300mm

By using slim and large temperature fixed points, as well as the modern ISO towers, standard thermometers of the ITS-90 (SPRTs as Pt100 or Pt25) and industrial reference thermometers (Pt100) can be calibrated from an immersion depth of 300 mm with excellent measurement uncertainties.

Comparison of SPRTs with IPRTs at temperature fixed points

When calibrating reference thermometers at temperature fixed points of the ITS-90, there is the question of the required calibration points. In order to make sure that the number of temperature fixed points of the ITS-90 is also enough in the higher temperature range, we have done extensive investigations. As a result of these investigations, the following graph shows the comparison of a standard thermometer (SPRT) with a reference thermometer.

The temperature difference shows that the maximum characteristic curve difference of approx. 25 mK is within the measurement uncertainty. For DAkkS calibrations in the higher temperature range, a plausibility check of the characteristic curve is done in addition to the mentioned tests. The test temperature is between the fixed point temperatures.

New digital archive Klasmeier Cloud:
Calibration documents quickly and easily available

The new Klasmeier Cloud – the perfect archive for all calibration documents. From now on, our customers will find a QR code on all devices calibrated by Klasmeier, with which they can quickly and easily access our digital archive.

Especially useful: In addition to all calibration documents, the contact details of the relevant contact person at Klasmeier are also always available there. So even five or more years from now, if you have any questions about your calibration product, you can find the information you need with just one click.

Finding the right calibration solution for you

Questions about our calibration services?


Please send your questions or enquiries to Mr Boris Kalb:

b.kalb@klasmeier.com
+49 661380 940-0